Ametek - nVision

Reference Pressure Recorder

Ametek - nVision

nVision is a multi-function pressure calibration device capable of recording measurement results...

Technical Description

  • The device is capable of measuring and recording pressure, temperature and current/voltage using independent modules that can be quickly replaced.
  • Can be used in the field and in the lab
  • Capable of recording up to 1 million data points
  • Pressure measuring module with wide range from vacuum to 1000 bar
  • Pressure accuracy to ± 0.0075% full range (below 30 bar)
  • Ability to measure differential pressure accurately to ± 0.025% of reading
  • Temperature measurement module using RTD sensor, accuracy up to ±(0.015% of reading + 0.02 Ω)
  • Module measures current up to 55 mA, voltage up to 28 VDC
  • Real-time graph digital display with anti-glare backlight
  • IP67 protection grade, dustproof, waterproof, chemical resistant, shockproof (tested at 5 feet height)
  • Compact device that can replace paper chart recorders
  • ATEX, IECEx, CE and DNV certified

Specifications

Main machine part (NV)
Display: 255 x 160 pixel graphic format, anti-glare backlit
Reading speed: 4 times/sec (standard mode); 10 times/sec (recording mode)
Power Supply: 60 days
Computer communication: mini-USB
Working temperature: -20 to 50 °C
Recorder: approx. 1 million points (in Single Recording mode)
Memory Type: non-volatile flash
Protection level: IP67 rating
Certification
ATEX II 1G Ex ia IIB T4 Ga hoặc T3 - SIRA 09ATEX2008X
IEC Ex ia IIB T4 Ga hoặc T3 - IECEx SIR09.0053X
CE and DNV
Weight: 680 g
Pressure Module (PM)
Pressure range: see table
Accuracy of conventional pressure gauge (for modules 3, 10 and 30 bar)
Range 0 to 30%: ± 0.0075% full range
Range 30 to 110%: ± 0.025% of reading
Vacuum range: ± 0.06% full range (-1.0 bar)
Accuracy of barometer type (BARO module 3 bar):
Range 0.0138 to 1.0000 barA: ± 0.0008 barA
Range 1.0000 to 4.0000 barA: ±(0.025% of reading) + 0.0003 barA
Pressure unit: bar, mbar, kPa, MPa, PSI, kg/cm2, inH2O, mmH2O, inHg, mmHg
Sensor materials (Wetted materials): 316SS or 316SS and Viton
Pressure connector: Crystal CPF female (CPF adaptor ra 1/4” NPT hoặc BSP hoặc M20)
Current and voltage module (MA20)
Current measurement: 
Measuring range: 
0 – 55 mA (MA20+)
0 – 25 mA (MA20)
Accuracy: ±(0.015% of reading + 0.002 mA)
Max. allowed current: 93.3 mA
Resolution: 0.001 mA or 0.01%
Unit: mA, % 4-20, % 10-50
Voltage measurement: 
Measuring range: 0 – 28 VDC
Accuracy: ±(0.015% of reading + 0.002 VDC)
Maximum allowable voltage: 30 VDC
Resolution: 0.001 VDC
Unit: VDC
Check the switch: 
Switch Type: dry contact
Closing resistor: < 10 Ω
Open resistor: > 10 MΩ
Temperature Module (RTD100)
Measuring range: 0 – 400 Ω (use with PRT/RTD 100 Ω)
Accuracy: ± (0.015% of reading + 0.02 Ω)
Resolution:0.01
Unit: °C, K, °F, R, Ω
Supported sensor types: 2, 3, 4 wires
Connector: M8 cable or terminal block

Related Products

413.jpg

Explosion-proof 2-channel pressure calibrator

Ametek - 30 series

412.jpg

2-channel pressure calibrator

Ametek - HPC40

411.jpg

2-channel pressure calibrator

Ametek - HPC50 Series

528.jpg

Medium range temperature calibration -45 to 155ºC

Ametek - RTC-156/157

Are you interested in our products?
Need a quote for a product or equipment?

Please contact our team of experts for free and professional advice