Technical Description
- Pressure calibration device with two ranges low and high up to 1000 bar
- Convenient for working at construction sites as well as inspection and calibration rooms.
- High accuracy up to ± 0.035% of reading
- Active temperature compensation technology ensures accuracy in readings without being affected by environmental conditions.
- Ability to measure and calibrate low pressure, high pressure and even vacuum pressure
- Calibration for mechanical pressure gauge, digital pressure gauge and transmitter
- Ability to connect to an external APM pressure module to change the measuring range
- Display on color LCD screen, 2 windows full of information
- Ability to connect external sensors to measure temperature
- Can simultaneously display the information value of both channels or the differential pressure value (value difference) of the two channels
- Sturdy design with protective rubber case, simple to use
Specifications
Overview | ||
1. Pressure measurement | ||
Scale | 3, 10 & 30 bar | -1 to 3, 10 & 30 bar |
100, 300, 700 & 1000 bar | 0 to 100, 300, 700 & 1000 bar | |
Accuracy level - | 0 to 30% of Range: | ±(0.01% of Full Scale) |
bar (Pressure gauge) | 30 to 110% of Range: | ±(0.035% of Reading) |
Vacuum* : | ±(0.05% of Full Scale**) | |
Unit | User optional | bar, mbar, kPa, MPa, psi, inH20, inHg, mmHg, mmH20, kg/cm2 |
2. Current and voltage measurement | ||
Current (mA)- Read | Accuracy level: | ±(0.015% of rdg + 0.002 mA) |
mA measuring range: | 0 to 55 mA | |
Resolution: | 0.001 mA or 0.01% | |
Unit: | mA and % | |
Current (mA)- Output | Accuracy level: | ± (0.015 of rdg + 0.002 mA) |
Measuring range: | 0 to 25 mA* | |
Voltage (VDC)- Read | Accuracy level: | ±(0.015 % of rdg + 2 mV) |
Measuring range: | 0 to 30 VDC | |
Resolution: | 0.001 VDC | |
Loop source | Fixed Output: | 24 VDC |
Check the switch | Type | Dry switch |
3. Temperature measurement | ||
Accuracy level: | ±(0.015% of rdg) + 0.02 Ohm | |
Measuring range: | 0 – 400 Ohms | |
Resolution: | 0.01 on all scales | |
Unit: | C, K, F, R, Ω | |
Connect: | Lemo Plug, 1S Series, 304 insert configuration | |
4. DISPLAY | ||
Screen | 320 x 240 pixel | |
Screen reading speed: | 3 times/paper (basic) | |
10 times/second (for switch test mode) | ||
5. Protection level | ||
Qualified | IP65 | |
6. Source | ||
Battery life | >12 hours in no power mode | |
>8 hours with 12mA power supply |