Medium range temperature calibration -22 to 155ºC
Ametek - RTC-158/250
Similar to RTC156/157 but RTC-158/250 has a special design, both dry tank and liquid tank in 1....
Gọi (+84) 828 31 08 99 để được tư vấn.
Technical Description
- Wide temperature range:
- RTC158: -30 to 155ºC
- RTC250: -45 to 155ºC
- Fully automatic calibration without computer connection
- Ideal thermal stability up to 0.005ºC
- Improved heat pipe technology with unique dual heat zone makes the temperature in the heat pipe synchronized
- Dynamic Temperature Compensation (DLC) technology allows the temperature in the heat pipe to be stable even when there are multiple sensors in the pipe (only available in models B and C)
- Color LCD screen, full information display (5 simultaneous temperature values; stable status; load compensation status; real-time clock; serial number of standard sensor; status of test sensor)
- Communicate with computer via USB port
- Smart standard sensor with “plug n play” plug designed at 90° angle for convenient plugging into heat pipe
- Switch test mode, Auto-stepping 20 steps easy to set up and use (B model only)
- Fast calibration, saving time
- The included JofraCal calibration software makes calibration and calibration certificate issuance simple and fast.
- RTD, TC, mA, V inputs (B model only)
- 24 VDC power supply for transmitter (B model only)
- Specially designed and convenient aluminum suitcase
Specifications
| Overview | |
| Temperature range (at ambient temperature 23°C): | |
| RTC158: | -22 to 155°C |
| RTC250: | 28 to 250°C |
| Accuracy: | |
| Using external standard STS sensor: | |
| RTC-158 B&C: | ± 0.04°C |
| RTC-250 B&C: | ± 0.07°C |
| Using internal standard sensor: | |
| RTC-158: | ± 0.18°C |
| RTC-250: | ± 0.28°C |
| Thermal stability: | |
| RTC-158: | ± 0.01°C |
| RTC-250: | ± 0.02°C |
| Thermal stabilization time: | Approximately 15 minutes |
| Resolution: | 1°, 0.1°, 0.01°, 0.001° |
| Heating rate: | |
| RTC-158: | -22 to 23°C 9 minutes |
| 23 to 155°C 51 minutes | |
| RTC-250: | 28 to 100°C 3 minutes |
| 100 to 250°C 9 minutes | |
| Cooling rate: | |
| RTC-158: | 155 to 23°C 33 minutes |
| 23 to -15°C 36 minutes | |
| RTC-250: | 250 to 100°C 24 minutes |
| 100 to 28°C 65 minutes | |
| Input for sensor to be tested (model B only): | |
| RTD: | Pt100, Pt500, Pt1000… 2, 3 or 4 wire type |
| TC: | Type E, J, K, T, R, S, N, U |
| Computer communication: | USB 2.0 |
| Power supply to transmitter: | 24VDC ± 10%, max. 28mA |
| mA transmitter input (B model only): | |
| Measuring range: | 0 to 24mA |
| Accuracy level: | ± 0.005%rdg + 0.01%FS |
| DC voltage measurement: | |
| Measuring range: | 0 to 12 VDC |
| Accuracy level: | ± 0.005%rdg + 0.01%FS |
| Thermowell Depth: | |
| Dry heat furnace version: | 180 mm |
| Liquid thermoelectric furnace version: | 150 mm |
| Power Supply: | 230V, 50Hz, 1150W max |
| Weight: | |
| RTC-158: | 28 kg |
| RTC-250: | 27 kg |