Medium range temperature calibration -45 to 155ºC
Ametek - RTCt-156/157
Building on the success of the RTC, the new RTCt offers significant improvements, representing a major leap forward in...
Gọi (+84) 828 31 08 99 để được tư vấn.
Technical Description
- Wide temperature range:
- RTCt-156: -32 to 155°C
- RTCt-157: -45 to 155°C
- Advanced touch screen interface, easy to operate
- Wi-Fi Connectivity – Remotely monitor, control and analyze calibration data via web browser interface.
- Dynamic Calibration Graph – Real-time stability monitoring for absolute accuracy.
- Dual Sensor Calibration – Increase efficiency with the ability to test two sensors at once. Fully automatic calibration without the need for a computer connection
- Ideal thermal stability up to 0.005ºC. Clearly identify drift, hysteresis and response delay
- Improved heat pipe technology with unique dual heat zone makes the temperature in the heat pipe synchronized
- Dynamic Temperature Compensation (DLC) technology allows the temperature in the heat pipe to be stable even when there are multiple sensors in the pipe (only available in models B and C)
- Color LCD screen, full information display (5 simultaneous temperature values; stable status; load compensation status; real-time clock; serial number of standard sensor; status of test sensor)
- Smart standard sensor with “plug n play” plug designed at 90° angle for convenient plugging into heat pipe
- Switch test mode, Auto-stepping 20 steps easy to set up and use (B model only)
- Fast calibration, saving time
- The included JofraCal calibration software makes calibration and calibration certificate issuance simple and fast.
- RTD, TC, mA, V inputs (B model only)
- 24 VDC power supply for transmitter (B model only)
- Specially designed and convenient aluminum suitcase
Applications
With their lightweight, compact design, the RTCt-156 and RTCt-157 are ideal for laboratory and field applications in the pharmaceutical, energy, aerospace and manufacturing industries.
Specifications
Overview | |
Temperature range: | |
RTCt-156: | -32 to 155°C |
RTCt-157: | -45 to 155°C |
Accuracy: | |
Using external standard sensor: | ± 0.04°C |
Using internal standard sensor: | ± 0.1°C |
Thermal stability: | max. ± 0.005oC |
Thermal stabilization time: | approximately 10 minutes |
Radial uniformity: | 0.01° |
Resolution: | 1°, 0.1°, 0.01° or 0.001° |
Heating rate: | |
RTCt-156: | -32 to 155°C 19 minutes |
RTCt-157: | -45 to 155°C 15 minutes |
Cooling rate: | |
RTCt-156: | 155 to 23°C 12 minutes |
23 to -32°C 25 minutes | |
RTCt-157: | 155 to 23°C 15 minutes |
23 to -45°C 42 minutes | |
Thermowell Depth: | 160 mm |
Input for standard STS sensor (4-wire RTD type) and DLC temperature compensated sensor (B & C models only) | |
Input for the sensor to be tested is RTD (2, 3 or 4 wire) and TC (B model only) | |
Power supply to transmitter: | 24 VDC ± 10% |
Transmitter input (B model only): | 0 to 24 mA with accuracy ± 0.005%rdg + 0.01%FS |
PC communication: | USB 2.0 port |
Power Supply: | 230V, 50Hz, max. 400W |
Weight: | 10.3 kg |