Fixed Real-Time X-ray System
GE - x|cube Series
Radiographic inspection equipment system used in casting industry, steel parts...optional 160 or 225kV generating tube power, sample can be tested...
Gọi (+84) 828 31 08 99 để được tư vấn.
Technical Description
- X-CUBE meets the needs for rapid testing of industrial samples such as: castings, steel, ceramics, plastics and special alloys.
- High voltage source 160kV or 225kV
- Test large samples up to 600mm in diameter and 900mm in height (XCube Compact model) or test up to 800mm in diameter and 1500mm in height (X-CUBE XL model)
- VISTAPLUS image analysis software helps evaluate and optimize inspection images
- Integrated CT imaging technology (optional)
Applications
- Used in mechanical manufacturing industry, special sample design in casting industry, large sample details
- Commonly used industries: automobiles, engine block manufacturing, casting industry...
- Fast defect detection, optional 3D imaging.
Specifications
Seifert x|cube | Compact | XL | ||
Energy (max.) | 160 kV | 225 kV | 160 kV | 225 kV |
Max. sample size (Ø x height in mm) | 600 x 900 ** | 600 x 900 ** | 800 x 1,500** | 800 x 1,500** |
Max. sample weight | 100 kg * | 100 kg * | 100 kg * | 100 kg * |
Cabinet dimensions, | 2,540 x 1,700 x 2,455 | 2,540 x 1,700 x 2,455 | 2,865 x 2,106 x 3,100 | 2,865 x 2,106 x 3,100 |
incl. backpack (L x W x H in mm) | ||||
Cabinet weight approx. | 3,950 kg | 4,550 kg | 6,550 kg | 6,550 kg |
Control panel weight approx | 350 kg | |||
Manipulation Travel | ||||
Max. horizontal motion | 650 mm | 850 mm | ||
across the X-ray beam | ||||
Max. external loading/ | 250 mm | |||
unloading position | ||||
Max. horizontal motion | 650 mm | 850 mm | ||
magnification axis | ||||
Focus detector distance (FDD) | 800-1,000mm | 1,000-1,200mm | ||
Max. vertical motion | 900 mm | 1,500 mm | ||
Max. tilt of the C arm | ± 45° | |||
Max. sample rotation | n x 360° | |||
2D software | Integrated image optimization system VISTAPLUS, optional with semi-automatic defect recognition (pass/fail assessment by the operator | |||
System control | Simple and intuitive system control and programming of recurring inspection tasks thanks to X-Touch® Panel | |||
Control / Drives | Hardware PLC PC independent / Fanuc servo drives | |||
Detector options | Selection of various digital image chains from the image intensifier to the flat panel detector. As an alternative to the digital image chain, e.g. temperature stabilized highly-dynamic GE DXR 250RT digital detector for pin sharp live images and very fast CT scans, the GE DXR 500 L detector for particularly high-resolution applications or the DXR 250 providing a large active area | |||
Flash!FiltersTM option | Proprietary live image optimization technology for easier visual defect detection in castings or weldings | |||
Tube options | Various mini to macro focuses as well as various high-power X-ray tubes up to 225 kV | |||
Computed tomography add-on | CT package contains all the required hardware and software components for combined 2D/3D operation with DXR 250 RT | |||
CT scan range | Max. 170mm Ø x 170mm height | |||
Min. voxel size | Up to 100µm, depending on the sample size | |||
Connection values / capacity | 3N PE 400/230V 50 Hz, 35 A, TN-S or TN-C-S network / approx. 11 kVA*** | |||
Earthing | Separate earthing for X-ray device and high-voltage generator (< 2 Ω) with at least 6 mm² | |||
Means of transport | Complete X-ray protection cabinet with fork lift truck / Control panel (on pallet) with fork lift truck | |||
Ambient conditions (in accordance with IEC 60 601-1) | Ambient temperature +10° C to +40° C, air pressure 700 hPa to 1060 hPa | |||
Compliant with national and international standards | ISO 9001; VDE 0100; UW; DIN EN 60204 (VDE 0113); VBG; DIN EN 60529 / IEC 529; German Radiation Control Act (RöV) of 1987 (with amendments in the current version); DIN EN 954-1; CFR 1020.40; DIN 54113 | |||
Radiation protection | The radiation safety cabinet is a full protective installation without type approval according to the German RöV. It complies with French NFC 74 100 and the US Performance Standard 21 CFR Subchapter J. For operation, other official licenses may be necessary. |