Heat exchange system testing equipment
GE - Apollo™
Eddy current heat exchanger tube tester with multi-channel and multi-frequency. Appolo is provided...
Gọi (+84) 828 31 08 99 để được tư vấn.
Technical Description
- Multi-channel, multi-frequency equipment
- Support probe types: RFT, ECT, ECA (eddy current array), air tube probe.
- 100% of data is processed in digital form, ensuring recording of all measurement data
- Configuration can be set up to 256 channels, expandable up to 1024 channels
- Support multiple testing methods simultaneously
- Wide frequency range from 1 Hz to 10 KHz with 32 frequency bands
- 8 probe signal inputs
- Automatic and manual gain control from 0-40dB
- Maximum sampling rate 50,000 samples/second
Applications
- Specialized in pipe inspection: external and internal defects with multiple channels and multiple frequency options
- Application of heat exchanger tube testing in power plants, air conditioners...
Specifications
| Overview | |
| Power Supply: | 100V to 240VAC, frequency 50 to 60Hz |
| Size: | 300 x 290 x 249mm, weight 8.7kg |
| Connection speed to PC: | 10/100Mbps |
| Signal encoder: | |
| Quantity: | 6 output ports (A,B,C) |
| Level: | LVTTL |
| Signal type: | A and B |
| Number of entries: | Number: 8 |
| Level: LVTTL | |
| Mode: Output Bit, Enable AQC, Tigger | |
| Number output: | Number: 8 |
| Level: LVTTL | |
| Modes: Output Bit, Alarm, AQC, Tigger, Time-Slot | |
| Width: programmable | |
| Analog input: | Number: 2 |
| Range: ±15V | |
| Bandwidth: 20kHz | |
| Analog output: | Number: 2 |
| Range: ±10V | |
| Bandwidth: 20kHz | |
| Eddy Current Mode (ECT): | |
| Access: | 8 |
| Channel: | 256 channels, expandable up to 1024 channels |
| Frequency number: | 32 |
| Frequency range: | from 1Hz to 10MHz |
| Transmitter: | 2 |
| Output port: | (2 direct transmission lines),(16 transmission lines) 100 Ohm |
| Generator Injection: | Mode Multiplex and Simultaneous |
| Operating potential: | 0 to 24V |
| Amplification: | Manual and automatic |
| Amplifier range: | 0 to 40 dB |
| Sampling rate: | 50,000 samples/second |
| Far-field eddy current (RFT) mode: | |
| Access: | 8 |
| Number of simultaneous channels: | 8 |
| Frequency: | 4 |
| Array mode (ECA): | |
| Access: | 8 |
| Number of simultaneous channels: | 256 channels, expandable up to 1024 channels |
| Frequency: | 32 |
| Frequency range: | from 1Hz to 10MHz |
| Transmitter: | 2 |
| Output port: | (2 direct transmission lines),(16 transmission lines) 100 Ohm |
| Generator Injection: | Mode Multiplex and Simultaneous |
| Operating potential: | 0 to 24V |
| Amplification: | Manual and automatic |
| Amplifier range: | 0 to 40 dB |
| Sampling rate: | 50000 samples/s |
| Probe balance: | Electric balance (push button) |
| Multiplexer: | Compatible with external multiplexers |
| Operating environment: | |
| Operating temperature: | 5°C → 40°C |
| Storage temperature: | -20°C → 70°C |
| Relative humidity | 90% |