Basic temperature calibration -10 to 400C
Ametek - ETC series
The ETC basic device line with the ability to calibrate infrared thermometers, also brings convenience when carrying to the field with a weight of less than 6kg
Gọi (+84) 828 31 08 99 để được tư vấn.
Technical Description
- Temperature range up to 400°C
- Fast calibration, saving time and cost
- Model ETC400R for infrared thermal gun standard
- Auto-stepping mode with 9 easy-to-use steps
- Fast heating and cooling time
- JofraCal calibration software can be used
- Compact design, sturdy aluminum suitcase
- Simple, easy to use
Specifications
Overview | |
Temperature range @ ambient temp. 23°C / 73°F | ETC-125 A |
Maximum.........................................................125°C / 257°F | |
Minimum @ ambient temp. 0°C / 32°F............ -18°C / -0°F | |
Minimum @ ambient temp. 23°C / 73°F............. -10°C /14°F | |
Minimum @ ambient temp. 40°C / 104°F............ 6°C / 43°F | |
ETC-400 A.......................28 to 400°C / 82 to 752°F@ 23°C | |
ETC-400 R.......................28 to 400°C / 82 to 752°F@ 23°C | |
Resolution (user-selectable) | Selectable ..............................................................1º or 0.1º |
Heating time | |
ETC-125 A | -10 to 23°C / 14 to 73°F..........................................3 minutes |
23 to 100°C / 73 to 212°F.....................................11 minutes | |
100 to 125°C / 212 to 257°F...................................7 minutes | |
ETC-400 A / R | 28 to 200°C / 82 to 392°F......................................2 minutes |
200 to 400°C / 392 to 752°F..................................3 minutes | |
Cooling time | |
ETC-125 A | 125 to 100°C / 257 to 212°F.................................. 1 minute |
100 to 0°C / 212 to 32°F.......................................17 minutes | |
0 to -10°C / 32 to 14°F..........................................14 minutes | |
ETC-400 A | 400 to 200°C / 752 to 392°F..................................6 minutes |
200 to 50°C / 392 to 122°F...................................15 minutes | |
ETC-400 R | 400 to 200°C / 752 to 392°F..................................9 minutes |
200 to 50°C / 392 to 122°F...................................24 minutes | |
Stability | ETC-125 A ...............................................±0.05°C / +0.09°F |
ETC-400 A .............................................. ±0.15°C / +0.27°F | |
ETC-400 R ................................................±0.3°C / ±0.54°F | |
Time to stability (approximate) | All models ..............................................................3 minutes |
Accuracy | ETC-125 A.................................................+0.5°C / +0.9°F 1) |
ETC-400 A................................................+0.5°C / +0.9°F 1) | |
ETC-400 R............................................... ±0.5°C / ±0.9°F 2) | |
ETC-400 R incl. emissivity.................................... ±0.4% rdg ±1°C / ±0.4% rdg. ±1.8°F |